An Accurate Isotope Dilution Liquid Chromatography-Tandem Mass Spectrometry Method for Serum C-Peptide and Its Use in Harmonization in China
نویسنده : Yuhang Deng, Chao Zhang , Jing Wang , Jie Zeng , Jiangtao Zhang Tianjiao Zhang , , Haijian Zhao , Weiyan Zhou , and Chuanbao
تاریخ انتشار : 1402/01/28
Methods: After protein precipitation with ZnSO4 solution, C-peptide was extracted from serum samples by anion-exchange solid-phase extraction and quantified by ID-LC–MS/MS
in positive ion mode. The precision and analytical recovery of the ID-LC–MS/MS method
were assessed. Seventy-six serum samples were analyzed using the ID-LC–MS/MS
method and six routine immunoassays. Ordinary linear regression (OLR) and Bland-Altman (BA) analyses were conducted to evaluate the relationship between the ID-LC–MS/
MS method and routine immunoassays. Five serum pool samples assigned using the IDLC–MS/MS method were used to recalibrate the routine assays. OLR and BA analyses
were re-conducted after recalibration.
Results: The within-run, between-run, and total precision for the ID-LC–MS/MS method at
four concentrations were 1.0%–2.1%, 0.6%–1.2%, and 1.3%–2.2%, respectively. The
analytical recoveries for the ID-LC–MS/MS method at three concentrations were 100.3%–
100.7%, 100.4%–101.0%, and 99.6%–100.7%. The developed method and the immunoassays were strongly correlated, with all R2
>0.98. The comparability among the immunoassays was substantially improved after recalibration.
Conclusions: The performance of the ID-LC–MS/MS method was carefully validated, and
this method can be used to improve the harmonization of serum C-peptide measurements
in China.